From packing an image plate to a finished X-ray on screen: an animated walkthrough, short lessons with Dürr Dental's own videos, cleaning instructions and a full fault-finder, all taken from the official manuals.
The VistaScan reads the X-ray stored on a reusable image plate, sends it to VistaSoft on your computer, then erases the plate ready for the next patient. This guide covers the VistaScan Mini Easy 2.0 with VistaSoft.
The whole routine in about a minute. Pause, step back, or click any step on the right to replay it.
Pack the image plate
Take the X-ray
Disinfect before unpacking
Feed the plate into the scanner
✓ Saved to the patient
This is a simplified illustration made by Dental Hygienics to guide you. Your scanner and the real VistaSoft screens look different.
Image plates are reusable several hundred times if you look after them. Most image problems start with how a plate was handled.
| Side | Looks like | Rule |
|---|---|---|
| Active | Light blue, with a positioning aid that shows on the X-ray | Always faces the X-ray tube. |
| Inactive | White, printed with back, the size and the maker's details | Visible through the cover. Faces you when you feed the scanner. |
The full routine, in the order the Dürr Dental manual gives it.
Finding your way around, working with patients, and viewing, improving and exporting images. Based on the VistaSoft 4.0 manual.
| Area | Where | What it is for |
|---|---|---|
| Menu bar | Top | Patients, image import, X-ray and video acquisition on the left. Reports, configuration and help on the right. |
| Patient data bar | Below the menu | Shows who is logged in, with the log-out button. |
| Image inspector | Left | The patient's image archive, grouped by date, newest first. |
| Light table | Centre | Where the images you select are displayed, side by side if you pick several. |
| Light table menu | Above the light table | Layouts, undo/redo, export, print, cloud upload, delete. |
| Toolbox | Right | Filters, corrections, measurements, drawings and annotations. |
| Status bar | Bottom | Messages, progress and notifications. |
A coloured bar beside each X-ray in the image archive shows whether the exposure was about right.
| Bar | Meaning |
|---|---|
| Green | Exposure is OK. |
| Yellow, high | Slightly too much. Reduce the dose a little. |
| Red, high | Too much. Reduce the dose. |
| Yellow, low | Slightly too little. Increase the dose a little. |
| Red, low | Too little. Increase the dose. |
Exposure settings are a matter for the dentist and your radiation protection arrangements. Ask Dental Hygienics if the bar is regularly off green.
Wear gloves, follow your local infection control rules, and read the instructions for each disinfectant. Dürr Dental has only tested the products named here.
| Item | How | Dürr Dental recommends | Never |
|---|---|---|---|
| Image plate | Wipe both sides with a soft, lint-free cloth before every use. Remove stubborn or dried-on dirt with the image plate cleaning wipe. Let it dry completely. | Dürr Dental image plate cleaning wipe | No disinfection wipes, no immersion, no steriliser. |
| Light protection cover | Disinfect before and after it goes in the mouth. Let it dry completely. | FD 333 forte wipes, FD 350, FD 322 premium wipes | Never reuse a cover. |
| Scanner surfaces | Remove the plate guide. Wipe off dirt with a soft, lint-free cloth dampened with cold tap water, then disinfect with a wipe. | FD 322, FD 333, FD 350, FD 366 sensitive | Never spray the unit or let liquid get inside. |
| Plate guide | Remove, wipe clean, disinfect with a wipe (or immersion disinfection), refit from above. | FD 322, FD 333, FD 350, FD 366 sensitive. Immersion: ID 212, ID 213 | No thermal disinfector or steam steriliser on any part of the scanner. |
| Image plate storage box and tray | Wipe clean, then disinfect. The storage tray may also go in a thermal disinfector or steam steriliser up to 134°C. | Box: FD 366 sensitive. Tray: FD 350, FD 366 sensitive | |
| Protective (dust) cover | Wipe with a damp, lint-free cloth. Only fit it to a clean, disinfected scanner. | Keep away from children (suffocation risk). |
Servicing is for trained engineers only. Dürr Dental's recommended schedule, based on about 15 intraoral images a day:
Find what you see on screen. Fixes are from the Dürr Dental manual. Tap a problem to open it.
What you can safely try yourself. For anything else, or if a code keeps coming back, note the number and contact Dental Hygienics.
| Code | Meaning | Try this |
|---|---|---|
| 10000 | Too much light on the unit | Darken the room, or turn the scanner away from direct light. |
| 10015 | Plate fed in off-centre | Feed the plate in centrally. |
| 10017 | Unit shutting down | Wait until it has shut down completely. |
| 1010 | Unit too hot | Let it cool down. Call us if it returns. |
| 1024, 1153, 1154, 2 | Internal communication or start-up fault | Switch the unit off, wait 10 seconds, switch back on. Call us if it returns (a firmware update may be needed). |
| 1026 | Incorrect acquisition mode | Choose a different acquisition mode in VistaSoft. Call us if it returns. |
| 1116 | Drive feed blocked | Call us. Do not force anything into or out of the slot. |
| 1118, 1121 | Input unit cover open, or fixing missing | Call us. This relates to the engineer's cleaning mode. |
| 1100, 1104, 1117, 1160, 1172, -1171 | Transport, erasure unit or laser fault | Stop using the scanner and call us. |
| -1008, 1022, 10009 | Internal fault, usually fixed by a firmware or software update | Call us to arrange the update. With 10009 the unit stays usable. |
| -78 | Storage medium full | Transfer the image data to the computer, or call us. |
| E2490 (in VistaSoft) | Connection to the scanner was interrupted | Restore the connection and repeat. |
From Dürr Dental's own YouTube channel. Filter by topic or search. Videos open on YouTube. Some show related scanner models or earlier VistaSoft versions, so details can differ.
No videos match. Try another word.
Dürr Dental's own documentation. Always follow the current instructions for your exact model.